Open original FDA record ↗Recall record
- FDA identifier
- Z-1371-2024
- Company in FDA record
- Hologic, Inc.
- FDA classification date
- Mar 22, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Use of assay along with plasma specimen may result in error code which would invalidate sample.
Source facts
- Action
- On February 23, 2024, Hologic issued a "Urgent Medical Device Recall Update" via FedEx. The update notified consignees of a sample dilution workflow for specific plasma specimens. On September 14, 2023 Hologic issued a "Urgent Medical Device Recall Notification" to affected consignees, via Fed Ex. Hologic provided the following information to customers: 1. Continue to use the Aptima CMV Quant assay. 2.When obtaining an ML2 error, please run the Mag Wash Clean Procedure and Contact Hologic Technical Support 3. With the Aptima CMV Quant assay, it is not recommended to retest a sample that was previously invalidated due to an ML2 error on the Panther instrument. 4. Complete Customer Response Form On (date) the firm issued a Customer Technical Bulletin, to consignees describing a sample dilution workflow for plasma specimens.
- Initiated
- 2023-09-14
- Root cause
- Use error
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, CT, FL, ID, IL, MD, MI, MO, TN, TX, WI and the countries of Australia, Germany, Denmark, Spain, France, Great Britain, Italy, Norway, Qatar, Sweden.
- Recall number
- Z-1371-2024
- Recall status
- Open, Classified
- CDRH record ID
- 206178
- FDA classified
- 2024-03-22
- Recall event number
- 94038