Z-1372-2019: Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Sch

Recall record Z-1372-2019 for Polymer Technology Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1372-2019
Company in FDA record
Polymer Technology Systems, Inc.
FDA record creation date
May 23, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being di P832 Chemistry (Diagnostic) CHH 2 Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412

Source facts

Action
The firm initially notified some of their customers from January through March 2019 with a Customer Bulletin. On March 28, 2019, the firm issued an updated Urgent Medical Device Recall letter to all affected customers. The letter alerted customers to the potential for test strips to slip out of the analyzer. Actions to be taken by the Customer/User: 1. Closely follow the instructions for proper testing. " Make sure the analyzer is sitting on a flat, stable surface. Do not perform testing with the analyzer held in the hand. " After inserting the test strip(s), do not move the analyzer until testing is complete and the test strip is removed. " As stated in the instructions for use, if an unexpected test result is given, the test should be repeated. There is no need to return or destroy the product. 2. Distributors should immediately forward this notice to all customers/users who may have received this product. 3. Acknowledge receipt of this instruction by completing the attached form, even if you do not have affected product on hand. If you have any questions, please contact Karmen Mercer, Director of Global Customer Care directly, Monday through Friday, 8:00 AM to 5:00 PM, Eastern Time at +1 317-870-5610 or (U.S. toll-free) 877-570-8610.
Initiated
2019-01-16
Root cause
Device Design
Terminated
2020-06-24
Distribution
Domestic distribution nationwide.
Recall number
Z-1372-2019
Recall status
Terminated
CDRH record ID
171370
FDA record created
2019-05-23
Recall event number
82331

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗