Open original FDA record ↗Recall record
- FDA identifier
- Z-1372-2022
- Company in FDA record
- Siemens Healthcare Diagnostics Inc
- FDA record creation date
- Jul 12, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Discrepant (low bias) pO2 results, could be an incorrect diagnosis of hypoxemia and initiation of unnecessary or inappropriate respiratory support.
Source facts
- Action
- Siemens has issued an Urgent Medical Device Correction letter POC 22-010.A.US via FedEx beginning on June 1, 2022. An Urgent Field Safety Notice (UFSN) was issued to OUS Siemens reporting sites with affected customers, for distribution and regional reporting to their local competent authorities. Letter states reason for recall, health risk and action to take: Please dispose any inventory of epoc BGEM test card lot 11-22005-50 currently in your possession in accordance with local and state disposal requirements. You may request free of charge replacement product from your local Siemens or distributor office. Please review your inventory of this product and assess your laboratory s replacement needs. Please use an epoc BGEM test card from another unaffected lot. Review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check and Product Replacement Form attached to this letter. Please retain this letter with your laboratory records and forward it to those who may have received this product. Questions contact your Siemens Customer Care Center or your local Siemens technical support representative.
- Initiated
- 2022-06-01
- Root cause
- Under Investigation by firm
- Distribution
- AL CA CO FL IA ID KS NM OK TX
- Recall number
- Z-1372-2022
- Recall status
- Open, Classified
- CDRH record ID
- 194382
- FDA record created
- 2022-07-12
- Recall event number
- 90432