Skip to content

Public FDA record

Z-1373-05: Baxter Healthcare Corporation, Medication Delivery Division, Syndeo PCA Syringe Pump, product codes 2L311

Recall record Z-1373-05 for Baxter Healthcare Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1373-05
Company in FDA record
Baxter Healthcare Corp.
FDA classification date
Aug 16, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

During an infusion in the 'Basal + PCA' or 'PCA only' modes, the SYNDEO pump may generate false Service Code 919 or Service Code 920, which will cause the pump to stop the infusion. Also, while attempting to turn on the device by pressing the On/Off key, the pump may not power up properly.

Source facts

Action
Baxter sent Important Product Information letters dated 7/6/05 to all Syndeo Pump customers, to the attention of the Director of Nursing, via first class mail. The letter informed the accounts of the scenarios they may experience regarding Service Codes 919 and 920, and the pump not being able to power up properly. Mitigation for Service Codes 919 and 920 is to turn the pump off, then turn it on again, and reprogram the infusion. If the codes occur again, the accounts were instructed to take the pump out of service and contact Baxter Medication Delivery Services at 1-800-843-7867. Any questions were referred to the Center for One Baxter at 1-800-422-9837. Baxter sent Product Withdrawal letters dated 12/11/09 to all Syndeo Pump customers, to the attention of the Director of Materials Management, with copies to the Director of Biomedical Engineering and the Director of Nursing. The letters informed the accounts that Baxter has stopped further development of the Syndeo pump platform, will pursue next generation PCA syringe pump technology and execute a controlled product withdrawal of the Syndeo pump over the next 12 months. Baxter sales representatives will contact the customers in the next few weeks to discuss options, including a Syndeo pump "buy back" plan, and to present alternative Baxter products to meet the customer's needs. The letters included a list of Syndeo pump serial numbers specific for each account, and requested the accounts to forward the letter to other facilities that they may have provided with Syndeo pumps. The accounts were requested to complete and fax back to Baxter the enclosed customer reply form, acknowledging receipt and understanding of the letter.
Initiated
2005-07-06
Root cause
Device Design
Terminated
2011-06-27
Distribution
Nationwide and Canada
Recall number
Z-1373-05
Recall status
Terminated
CDRH record ID
40398
FDA classified
2005-08-16
Recall event number
32589

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗