Skip to content

Public FDA record

Z-1373-2013: EZ Breathe Medication Cup. For the delivery of liquid medications for respiratory usage.

Recall record Z-1373-2013 for Nephron Pharmaceuticals Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1373-2013
Company in FDA record
Nephron Pharmaceuticals Corp.
FDA classification date
May 29, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Possible dislodgement of the "Plate A", if this instance occurs, it presents a potential choking hazard.

Source facts

Action
The firm, Nephron Pharmaceuticals Corporation, notified its consignees on April 30, 2013, and May 3, 2013, of the problem via a recall notice and a Press Release, respectively. The firm directed consignees to discontinue use of any EZ Breathe atomizer units contained in the Asthmanefrin Starter Kits with the affected lot numbers and also those that were sold individually. Customers were directed to return all affected product.
Initiated
2013-04-30
Root cause
Nonconforming Material/Component
Terminated
2016-06-28
Distribution
Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AL, AZ, CA, CO, CT, FL, IL, IN, IO, GA, KS, KY, LA, MD, MN, MS, MO, NK, NV, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN , TX, VT, VA, WA, and WV, and the countries of Guatemala, Poland, Panama, and United Kingdom.
Recall number
Z-1373-2013
Recall status
Terminated
CDRH record ID
117836
FDA classified
2013-05-29
Recall event number
65070

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗