Recall record Z-1373-2024 for Ortho-Clinical Diagnostics. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1373-2024
Company in FDA record
Ortho-Clinical Diagnostics
FDA classification date
Mar 22, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could lead to testing delays and erroneous patient results. This issue caused positive or negative bias in the quantification of sodium (Na+) and total carbon dioxide (ECO2). The incorrect diluent into the Performance Verifier II exhibited a potential negative bias of up to 6 mmol/L for ECO2 and Na+ testing.

Source facts

Action
Consignees were sent an URGENT PRODUCT CORRECTION NOTIFICATION dated 2/27/24 informing them that for product Lot Q1174, diluent Lot M9815 was found in some sales units instead of the expected diluent Lot Q1175, and for product Lot R1176, diluent Lot P9967 was found in some sales units instead of the expected diluent Lot R1177. Consignees were asked to inspect their inventory to see if any mismatched product from affected lots was present and to discard any affected units. The provided Confirmation of Receipt form is to be returned by 3/8/24. Results received using affected devices are to be reviewed for any unexpected QC shift. The recall notification is to be forwarded if any devices were further distributed and the notification maintained in the User Documentation. Customers with any questions can call Global Services at 1-800-421-3311.
Initiated
2024-02-27
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
Recall number
Z-1373-2024
Recall status
Open, Classified
CDRH record ID
206519
FDA classified
2024-03-22
Recall event number
94174

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗