Philips Medical Systems North America Co. Phillips
FDA classification date
Sep 16, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
Unexpected movement of the X-ray table may occur. These movements may lead to patient injury, e.g., an uncontrolled tilt may cause the patient to slip off.
Source facts
Action
On 3/22/08, the firm mailed letters to their customers via certified mail. The letter informed the customers of the recall, advising the customers to, "Use the emergency stop button or release the initiated movement command to halt undesired table movement. If you perform digital and conventional acquisition, you can avoid the problem by closing the single examinations daily or by clearing the patient database regularly." The letter further informed the customers that a Philips representative will be installing updated software at a future date.
Contact Sarah Baxter at 425-487-7665 or Philips Call Center at 800-722-9377, #5, #2 if you have questions.
Initiated
2008-03-22
Root cause
Software design
Terminated
2010-09-17
Distribution
Nationwide Distribution -- including states of CA, CO, KY, TN, LA, MI, FL, MS, TX, VA, WI, MO, AZ, and NY.
Recall number
Z-1374-2008
Recall status
Terminated
CDRH record ID
69577
FDA classified
2008-09-16
Recall event number
47489