Skip to content

Public FDA record

Z-1374-2018: Fornier Phantom Fiber(TM) Sutures: Item Number Description SMB000425 4.5 MM PHANTOM FT BIOCOMPOSITE SUTUR

Recall record Z-1374-2018 for Tornier, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1374-2018
Company in FDA record
Tornier, Inc
FDA record creation date
Apr 13, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Data has shown that degradation of a raw material used in Phantom Fiber may occur more quickly than expected if improperly stored. This may impact the shelf-life of the product which could cause the suture to lose strength more quickly than expected.

Source facts

Action
The firm, Wright, sent an "Urgent Field Safety Notice" letter by email on 02/08/2018. The firm followed up with a letter on April 02, 2018 disseminated by FedEx Priority mail to their customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following actions: " Cease using the affected devices immediately. " Medical facilities are to determine if any of the recalled implants are on hand, and return affected implants immediately to Wright Medical. "Review the notice and complete the Acknowledgement Section Attachment C to signify that your facility has been informed of this Field Safety Notice (Recall). " Return the completed Acknowledgement to Wright Medical via email to: FieldAction@wright.com within five (5) working days of this notice. "Contact customer service at 1-888-494-7950 to facilitate any product returns. For any inquiries regarding the Phantom Fiber Field Safety Notice (Recall) contact: Wright Medical Quality Systems Manager by Email: FieldAction@wright.com or Phone: 952.683.7482.
Initiated
2018-02-08
Root cause
Nonconforming Material/Component
Terminated
2020-08-07
Distribution
US Nationwide Distribution.
Recall number
Z-1374-2018
Recall status
Terminated
CDRH record ID
162146
FDA record created
2018-04-13
Recall event number
79373

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗