Z-1374-2021: Arrow OnControl Bone Lesion Biopsy System Tray-11 ga 10 cm Access/13 ga 14 cm Biopsy, Product Code: 9464-

Recall record Z-1374-2021 for Arrow International Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1374-2021
Company in FDA record
Arrow International Inc
FDA classification date
Apr 7, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Difficulty inserting the biopsy ejector rod into the bone lesion biopsy cannula, preventing the biopsy sample from being collected. Use of a defective device is likely to require immediate repeat biopsy, leading to a prolonged procedure

Source facts

Action
Teleflex issued to US consignee notification letter on March 8, 2021 via FedEx 2-day mail. Letter states reason for recall, health risk, and action to take: Action list number 1 Medical facilities 1. We request that you check your inventory for product within the scope of this FSCA. Users should cease use and distribution of impacted product and quarantine immediately. 2. If you do not have stock in scope of this FSCA mark the according checkbox on the Acknowledgement Form (Appendix 1) and return the form to the fax number or e-Mail address mentioned below. 3. If you do have stock in scope of this FSCA, mark the according checkbox on the Acknowledgement Form (Appendix 1) and contact customer service by calling the phone number mentioned below. Customer service will issue you with a return number. Write the return number into the respective field in the Acknowledgement Form and return this form immediately to Customer Service. 4. Teleflex (or your local dealer) will issue a credit note upon receipt of the returned affected product. Action list number 2 Distributors 1. Provide this field safety notice to all customers who have received product in scope of this FSCA. Your customer is then required to complete the acknowledgement form and return to you. 2. We request that you check your inventory for product within the scope of this FSCA. Cease use and distribution of impacted product and quarantine immediately. You may then return all product in scope. 3. As a distributor, you are then required to confirm to Teleflex that you have completed the field activity outlined above. Upon completion of your actions, please forward the completed Acknowledgement Form to Customer Service. 4. If you have further distributed product outside of your country, please notify Teleflex by return email to the e-Mail address below. Should you require any further information or support concerning this issue, please contact: Customer Service: Contact: Customer Serv
Initiated
2021-03-08
Root cause
Process design
Distribution
US Nationwide distribution.
Recall number
Z-1374-2021
Recall status
Open, Classified
CDRH record ID
186193
FDA classified
2021-04-07
Recall event number
87499

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗