Z-1374-2022: Ventana HE 600 System, automated slide preparer, for use in laboratories.

Recall record Z-1374-2022 for Ventana Medical Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1374-2022
Company in FDA record
Ventana Medical Systems Inc
FDA classification date
Jul 12, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for fluid leak inside the Ventana HE 600 instrument which could result in electrical short and fire

Source facts

Action
On June 6, 2022 Ventana Medical Systems, Inc. (via Roche Diagnostics Corp.) issues an "Urgent Medical Device Correction" notification to affected consignees. Ventana Medical ask consignees to take the following actions: 1. Notify all operators of this potential hazard and forward this UMDC as appropriate. If your facility has distributed the affected device to another site, please ensure this UMDC is provided to that site. 2. Do not leave the instrument unattended when the power is on. Turn off the power to the instrument when it is unattended per the instructions in this UMDC. 3. Do not place the system in stand-by mode overnight when unattended until further mitigation actions have been implemented. 4. If fire is detected, follow your local emergency guidelines. 5. If a leak, smoke, or odor is detected in your VENTANA HE 600 instrument, power off the instrument immediately and call the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-227-2155. 6. Complete the enclosed faxback form (TP-01573) and return it according to the instructions on the form. 7. File this Urgent Medical Device Correction (UMDC) for future reference. 8. Adverse events or quality problems experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Events Reporting Program: Online at http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail), or call FDA 1-800-FDA-1088.
Initiated
2022-06-01
Root cause
Under Investigation by firm
Distribution
US: AL AR AZ CA CO CT DC FL GA HI IA IL IN KS KY MA MD MI MN MO MS MT NC NE NJ NY OH OK OR PA PR SC SD TN TX UT VA WA WI WV OUS: Canada, Austria, Belgium, Croatia, Denmark, Finland, France, Germany, Greece, Ireland, Italy, Luxembourg, Netherlands, Norway, Poland, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, UK, Argentina, Australia, Brazil, Chile, China, Colombia, Costa Rica, Ecuador, Egypt, El Salvador, Guatemala, Hong Kong, Israel, Japan, Kuwait, Lebanon, Malaysia, Saudi Arabia, Singapore, South Korea, Taiwan, Thailand, United Arab Emirates (UAE), Vietnam
Recall number
Z-1374-2022
Recall status
Open, Classified
CDRH record ID
194409
FDA classified
2022-07-12
Recall event number
90363

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗