Open original FDA record ↗Recall record
- FDA identifier
- Z-1374-2024
- Company in FDA record
- Boston Scientific Corporation
- FDA classification date
- Mar 27, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
An investigation determined that delivery of the Obsidio embolic using the aliquot technique for lower gastrointestinal bleeding embolization poses a high risk of bowel ischemia. The most serious and the most common adverse health consequence, reasonably foreseeable to occur, is the need to perform major surgery such as bowel resection and/or diverting colostomy. Therefore, Boston Scientific does not recommend that the aliquot technique be used to deliver the Obsidio device for lower GI bleed embolization procedures.
Source facts
- Action
- Boston Scientific issued an URGENT MEDICAL DEVICE PRODUCT ADVISORY notice to its consignees via FedEx on 02/21/2024. The notice explained the use of the aliquot technique in not recommended for lower GI bleed embolization procedure due to the risk to the patient. The firm directed the consignee to immediately post the notice near the product to ensure the information is easily accessible to all users. For questions, please contact your local Boston Scientific Representative. Updated recall strategy: All Obsidio accounts were sent an updated Product Advisory, which includes the Customer Cover Letter, Customer Advisory Letter and Sample RVTF, to inform physicians of updated IFU warnings and instruction regarding delivery techniques. The customer notifications were sent on Friday, October 11, 2024, via delivery using Fed Ex, DHL or other equivalent service. Accounts are not asked to return affected Obsidio product.
- Initiated
- 2024-02-20
- Root cause
- Under Investigation by firm
- Distribution
- US
- Recall number
- Z-1374-2024
- Recall status
- Open, Classified
- CDRH record ID
- 206388
- FDA classified
- 2024-03-27
- Recall event number
- 94126