Z-1374-2024: Obsidio Conformable Embolic, Material Numbers (UPN): a) M0013972001010, b) M0013972101010

Recall record Z-1374-2024 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1374-2024
Company in FDA record
Boston Scientific Corporation
FDA classification date
Mar 27, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

An investigation determined that delivery of the Obsidio embolic using the aliquot technique for lower gastrointestinal bleeding embolization poses a high risk of bowel ischemia. The most serious and the most common adverse health consequence, reasonably foreseeable to occur, is the need to perform major surgery such as bowel resection and/or diverting colostomy. Therefore, Boston Scientific does not recommend that the aliquot technique be used to deliver the Obsidio device for lower GI bleed embolization procedures.

Source facts

Action
Boston Scientific issued an URGENT MEDICAL DEVICE PRODUCT ADVISORY notice to its consignees via FedEx on 02/21/2024. The notice explained the use of the aliquot technique in not recommended for lower GI bleed embolization procedure due to the risk to the patient. The firm directed the consignee to immediately post the notice near the product to ensure the information is easily accessible to all users. For questions, please contact your local Boston Scientific Representative. Updated recall strategy: All Obsidio accounts were sent an updated Product Advisory, which includes the Customer Cover Letter, Customer Advisory Letter and Sample RVTF, to inform physicians of updated IFU warnings and instruction regarding delivery techniques. The customer notifications were sent on Friday, October 11, 2024, via delivery using Fed Ex, DHL or other equivalent service. Accounts are not asked to return affected Obsidio product.
Initiated
2024-02-20
Root cause
Under Investigation by firm
Distribution
US
Recall number
Z-1374-2024
Recall status
Open, Classified
CDRH record ID
206388
FDA classified
2024-03-27
Recall event number
94126

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗