Open original FDA record ↗Recall record
- FDA identifier
- Z-1375-2021
- Company in FDA record
- Pacific Medical Group Inc.
- FDA record creation date
- Apr 19, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Aftermarket front bezel components were installed during service/repair, not by the original manufacturer, within infusion pump modules. Over time, the posts on the front bezel component may crack or separate. Separation of one or more bezel posts could lead to free flow, over infusion, under infusion, or interruption of infusion.
Source facts
- Action
- On 03/24/2021, Medical Device Recall Notifications were emailed to customers. Customers were asked to take the following actions: If you have affected product, remove these devices from clinical use and immediately quarantine them until the impacted parts can be returned to the recalling firm; contact Customer Service 800-449-5328, pm.recallapm@avantehs.com. The recall part number is RCL2021-823-01. Once returned, the affected device will be evaluated and inspected and where the component under recall is identified, it will be replaced by our organization with the OEM component and then returned. If your facility has used the component under recall in repairs of the top assembly immediately stop use and quarantine the affected product until the impacted parts are replaced. If you have resold/distributed the affected product, please forward this notice to your customers. Complete and submit the recall acknowledgment form to pm.recallapm@avantehs.com
- Initiated
- 2021-03-24
- Root cause
- Component design/selection
- Distribution
- Distribution US nationwide.
- Recall number
- Z-1375-2021
- Recall status
- Open, Classified
- CDRH record ID
- 186149
- FDA record created
- 2021-04-19
- Recall event number
- 87483