Open original FDA record ↗Recall record
- FDA identifier
- Z-1375-2022
- Company in FDA record
- DePuy Orthopaedics, Inc.
- FDA record creation date
- Jul 12, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The Cathcart Ball has an inherent +5mm offset which is not currently specified in its Instructions for Use or Surgical Technique Guide. The STG has been updated to reflect the clarification that the +5mm offset should be taken into consideration when using a Cathcart head in a construct for hemi-hip arthroplasty procedures.
Source facts
- Action
- On June 13, 2022, the firm notified distributed recall letters to customers titled URGENT FIELD SAFETY NOTICE. Customers were instructed to contact their local sales consultant to receive a copy of the revised STG (reference document number DSUS/JRC/0317/2044 Rev B). Product is not being removed from the field.
- Initiated
- 2022-05-18
- Root cause
- Labeling design
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Canada, Denmark, Finland, Germany, Greece, Guam, Hong Kong, Ireland, Netherlands, New Zealand, Norway, South Africa, Spain, Sweden, Switzerland, United Kingdom.
- Recall number
- Z-1375-2022
- Recall status
- Open, Classified
- CDRH record ID
- 194540
- FDA record created
- 2022-07-12
- Recall event number
- 90476