Open original FDA record ↗Recall record
- FDA identifier
- Z-1375-2025
- Company in FDA record
- CareFusion 303, Inc.
- FDA classification date
- Mar 17, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Sync timeout may occur between the server and automated dispensing device, due to a server software error which may occur if syncing a large amount of data that takes excessive time, which may cause the dispensing devices to become disconnected from the ES server; dispensing devices will not receive updated information, which may lead to incorrect therapy, and delay in accessing medication.
Source facts
- Action
- On 2/6/2025, correction notices were mailed to customers informing them of the following: 1) Firm will be contacting customers and distributors to schedule time to upgrade software to 1.7.4.138 or a later release. 2) Identify all customers within your distribution network that purchased any affected product as defined in this notification. Provide a copy of the attached customer letter to all customers to advise them of this field action notification. 3) Complete and return the customer/distributor response form via email to BDRC8@bd.com Report complaints to firm via the North American Regional Complaint Center: Phone: 1-844-8BD-LIFE (1-844-823-5433) Say "Product Complaints" when prompted Mon-Fri 8:00am - 5:00pm CT Email: productcomplaints@bd.com Recall/Remediation Related Questions: 1-866-583-8783, 8:00am ET to 7:00pm ET), Monday - Friday. Product Complaints or Technical Support: 1-800-727-6102 Phone hours: 24 hours, 7 days a week Website: bd.com/self-service
- Initiated
- 2025-02-06
- Root cause
- Software design
- Distribution
- Worldwide - US Nationwide distribution including in the states of KY, OH, NY, MO, NC, PA, FL, TN, CT, LA, MN, VA, MD, CA, TX, NE, IA, OK, MI, PR, SD, IN, ME, KS, IL, NM, WA, AR, HI, MS, GA, WY, WI, VT, AL, NV, NJ, DC, AZ, MT, AK, OR, CO, NH, MA, SC, ID, UT, ND and the countries of TW, CA, AE, MX, SG, AU, NZ, GB, CH, BH.
- Recall number
- Z-1375-2025
- Recall status
- Open, Classified
- CDRH record ID
- 212311
- FDA classified
- 2025-03-17
- Recall event number
- 96202