Open original FDA record ↗Recall record
- FDA identifier
- Z-1376-2008
- Company in FDA record
- Biomet, Inc.
- FDA classification date
- Jul 25, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Mislabeled: The packaging of the two products was mixed up, resulting in the products being labeled with an incorrect size. If implanted, the product may not fit properly, which could require repeat surgery. The problem could also result in a delay in the procedure.
Source facts
- Action
- Distributors were notified via recall letter (Urgent Medical Device Removal Notice) dated 2/12/08 to locate the devices in their possession and at hospital sites, discontinue use, and return them to Biomet. Hospitals having received the devices will be provided with a recall letter dated 2/12/08 which describes the nature of the problem and the removal of these instruments. A third letter dated 2/12/08 was provided for implanting physicians.
- Initiated
- 2008-02-12
- Root cause
- Packaging process control
- Terminated
- 2009-12-22
- Distribution
- Worldwide: USA and Australia
- Recall number
- Z-1376-2008
- Recall status
- Terminated
- CDRH record ID
- 67991
- FDA classified
- 2008-07-25
- Recall event number
- 46708