Skip to content

Public FDA record

Z-1376-2022: ABACUS TPN (Total Parenteral Nutrition) Calculation software, Product codes: 8300-0167 (Abacus V3.1 CE),

Recall record Z-1376-2022 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1376-2022
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Jul 22, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

There is a potential risk of medication error when using specific Abacus Order Entry & Calculation software.

Source facts

Action
Baxter sent an Urgent Medical Device Correction notice on 06/22/2022 by letter. The notice explained the problem and the hazard and recommended the following: customers refrain from making updates to bag label templates and to contact Baxter Technical Support if updates to bag label templates are needed. Please continue to follow the instructions in the Abacus User and Abacus Configuration Guides when creating an order, and have a pharmacist thoroughly review all order outputs including the printed label for accuracy and safety. As mentioned above, please ensure your facility processes include pharmacy checks during the compounding process as well as a nursing check. Baxter will communicate to customers when the software upgrade becomes available. At that time additional instructions to obtain the software upgrade will be provided.
Initiated
2022-06-22
Root cause
Software design
Terminated
2024-10-09
Distribution
United States (including Puerto Rico), Argentina, Bermuda, Chile, Colombia, Dominican Republic, Panama, Canada, Vietnam, Hong Kong, Singapore, and EMEA
Recall number
Z-1376-2022
Recall status
Terminated
CDRH record ID
193835
FDA classified
2022-07-22
Recall event number
90212

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗