Skip to content

Public FDA record

Z-1376-2023: Digital VIPR ROC-9839 Roberts Oxygen Ready to Use Digital VIPR (head only) ROC-9840, Roberts Oxygen Read

Recall record Z-1376-2023 for Western/Scott Fetzer Company. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1376-2023
Company in FDA record
Western/Scott Fetzer Company
FDA classification date
Apr 12, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Gauge malfunction: Digital gauge displays an incorrect tank pressure - displays pressure higher than actual tank pressure. This may lead to tank running out of oxygen during use without audible alarms, potentially leading to an out of oxygen event and no flow.

Source facts

Action
On January 3, 2023, the firm issued a Safety Alert to customers. Customers were provided with updated operating instructions. Customers instructed to take the following immediate actions: 1. Implement updated filling and venting instructions which include steps for the filler to verify proper gauge pressure after vent down and after fill/re-fill. 2. Implement an additional verification of gauge pressure upon delivery at the healthcare facility. 3. Take care to minimize any impact to the cylinder during use, transportation, or processing during fill/re-fill. The firm continues to evaluate the gauge error issue and will implement additional corrective actions as appropriate.
Initiated
2023-01-03
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of CO, CT, ID, MD, MT, TN, WA.
Recall number
Z-1376-2023
Recall status
Open, Classified
CDRH record ID
198183
FDA classified
2023-04-12
Recall event number
91561

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗