Z-1377-05: Cytomics FC500 MPL with MXP Software versions 1.1 and 2.0

Recall record Z-1377-05 for Beckman Coulter Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1377-05
Company in FDA record
Beckman Coulter Inc
FDA classification date
Aug 17, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Software anomally. If the laser is shutdown or fluctuates outside of the specification limits, the software continues to run and will not indicate a laser failure, and could therefore cause erroneous results.

Source facts

Action
Letters issued to users on July 13, 2005 which also included modified operating instructions, and response cards.
Initiated
2005-07-06
Root cause
Other
Terminated
2012-04-26
Distribution
USA
Recall number
Z-1377-05
Recall status
Terminated
CDRH record ID
40707
FDA classified
2005-08-17
Recall event number
32733

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗