Z-1377-06: Accu-Chek Comfort Curve blood glucose test strips; (packaged in 10s, 50s, and/or 100s) U.S. Reference nos

Recall record Z-1377-06 for Roche Diagnostics Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1377-06
Company in FDA record
Roche Diagnostics Corp.
FDA classification date
Aug 11, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The drying agent beads, located in the cap, may become loose and fall into the vial causing the test strips to potentially give incorrect blood glucose test results.

Source facts

Action
U.S. consignees, point of care coordinators or laboratory managers, long term care professionals, health care professionals, pharmacists, and mail order distributors were notified via letter dated 6/28/06 to cease using the strips and to contact the firm for replacement if they see loose beads in their vial. International accounts were also notified.
Initiated
2006-06-27
Root cause
Other
Terminated
2007-05-04
Distribution
Worldwide Distribution-USA and countries of Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Ecuador, England, France, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Mexico, New Zealand, Pakistan, Paraguay, Peru, Philippines, Singapore, South Africa, South Korea, Spain, Sweden, Taiwan, Thailand and Uruguay.
Recall number
Z-1377-06
Recall status
Terminated
CDRH record ID
46695
FDA classified
2006-08-11
Recall event number
35729

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗