Skip to content

Public FDA record

Z-1377-2021: BD BBL Martin-Lewis Agar, 100 Ea - Product Usage: used for the isolation of pathogenic Neisseria from sp

Recall record Z-1377-2021 for Becton Dickinson & Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1377-2021
Company in FDA record
Becton Dickinson & Co.
FDA record creation date
Apr 8, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Products show heavy growth of C. albicans after 24 hours of incubation due to misformulation (expected reaction is partial inhibition after 18-48 hours incubation) and reduce the identification of Neisseria contributing to a delay in time to definitive diagnosis or a delay in proper treatment

Source facts

Action
BD issued Urgent Medical Device Correction letter on 3/17/21 to customers and distributors via FedEx tracked letter. Letter states reason for recall, health risk and action to take: 1.Immediately inspect your inventory for the specific catalog and lot numbers listed above. Destroy all products subject to the recall following your institutions process for destruction. 2.Share this recall notification with all users within your facility network to ensure they are also aware of this recall. 3.Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your product replacement, if required. BD will provide a product replacement for all destroyed inventory. If you require further assistance, please contact: BD Contact US Contact Information Outside of US BD Customer/ Technical Support 800-638-8663 Monday Friday 7:00am and 7:00pm EST For customers outside the US, contact the local BD representative or distributor.
Initiated
2021-03-17
Root cause
Under Investigation by firm
Terminated
2022-08-05
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, Korea.
Recall number
Z-1377-2021
Recall status
Terminated
CDRH record ID
186259
FDA record created
2021-04-08
Recall event number
87546

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗