The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.
Source facts
Action
Cook plans to begin notifying consignees of the recall on January 12, 2018
Written recall communication letters will be mailed via courier to US consignees
Initiated
2018-01-12
Root cause
Device Design
Terminated
2020-05-08
Distribution
Nationwide (K-IMNC-100 -US G46022) . (K-IMNC-100 G20079 manufactured in Australia and distributed worldwide)
Recall number
Z-1378-2018
Recall status
Terminated
CDRH record ID
163147
FDA classified
2018-04-13
Recall event number
79711