Z-1378-2023: BioPlex 2200, Anti-CCP Calibrator Set, REF 663-3200, IVD

Recall record Z-1378-2023 for Bio-Rad Laboratories, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1378-2023
Company in FDA record
Bio-Rad Laboratories, Inc.
FDA classification date
Apr 12, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positivesBio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positivesBio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positives.There is no performance issue with the calibrator lot# 53727. They are included in the recall only because the calibrators are lot matched to the impacted reagent pack.

Source facts

Action
On February 21, 2023, Bio-Rad Laboratories issued a "Urgent: Medical Device Recall" notification to affected consignees. In addition to informing consignees about the recall, Bio-Rad asked consignees to take the following actions: 1. We are requesting that customers discontinue the use of and dispose of according to local waste management procedures any unused BioPlex 2200 Anti-CCP Reagent Pack (Catalog No. 665-3250, UDI GTIN 00847865000857) Lot No. 301481 and BioPlex 2200 Anti-CCPCalibrator Set (Catalog No. 663-3200, UDI GTIN 00847865000819) Lot No. 53727. 2. If you have distributed or transferred this product to any other Bio-Rad customers, or to any other laboratories from your site, please notify those customers of this field action. 3. Please complete the attached Customer Response Form and return it to your local Bio-Rad Technical Support so they can assist you with obtaining replacement reagent packs and calibrator sets Bio-Rad is working to provide a new lot of BioPlex 2200 Anti-CCP Reagent Pack and matching calibrator sets as soon as possible. Please provide the number of replacement units required in the attached customer response form.
Initiated
2023-02-21
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Czech Republic, Germany, Spain, France, United Kingdom, Israel, Italy, Norway, Sweden.
Recall number
Z-1378-2023
Recall status
Open, Classified
CDRH record ID
199062
FDA classified
2023-04-12
Recall event number
91854

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗