Skip to content

Public FDA record

Z-1379-2021: Trial Cutting Guide for Triathlon Total Knee System, Triathlon PKR System, Restoris MCK Knee System, pros

Recall record Z-1379-2021 for Howmedica Osteonics Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1379-2021
Company in FDA record
Howmedica Osteonics Corp.
FDA classification date
Apr 9, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The posterior condyle of the Specialty Triathlon TS Trial Cutting Guide(s), sizes 1 through 8, (Left and Right) has the potential to fracture at the augment cut slot. This can occur in the Trial Cutting Guide packaging during transport or during impaction of the instrument in surgery.

Source facts

Action
On 3/4/2021, Stryker issued an Urgent Medical Device Recall notice to customers via letter due to Stryker has initiated a voluntary, catalog number specific recall for the Specialty Triathlon TS Trial Cutting Guide, sizes 1 through 8, (Left and Right). The intent of this letter is to list hazards and harms potentially associated with the above referenced product and to list any risk mitigation factors. For questions or concerns after reviewing this letter, please contact Customer Service at (201) 831-5000.
Initiated
2021-03-05
Root cause
Under Investigation by firm
Terminated
2024-05-10
Distribution
Distribution in US to: CA,, GA, NC, NJ, NV, OK, TX, and to Canada
Recall number
Z-1379-2021
Recall status
Terminated
CDRH record ID
186218
FDA classified
2021-04-09
Recall event number
87517

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.