Z-1380-05: Accura System for Blood Filtration, product codes 5M5660 and ACCURA01 (international); a hemofiltration s

Recall record Z-1380-05 for Baxter Healthcare Renal Div. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1380-05
Company in FDA record
Baxter Healthcare Renal Div
FDA classification date
Aug 18, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

In the event therapy is terminated, as part of routine instrument programming or therapy maintenance, the Accura instrument may experience difficulty in performing a successful self-test (failure) upon an immediate attempt to restart therapy

Source facts

Action
Baxter sent Important Product Information letters dated 7/13/05 to all Accura device customers, to the attention of the Director of Nursing, via first class mail. The letter informed the accounts of the on-going investigation into the potential self-test failure associated with the Accura instrument, and provided the customers with steps to follow if they encounter failed instrument self-tests. These guidance steps include: 1) Place the instrument into 'stand-by' mode. Generally, 15 minutes in 'stand-by' mode is sufficient to allow successful instrument self-test procedures. Enter 'stand-by' mode by turning off the power by using the upper power switch, or by selecting 'Accura off' from the user interface. Avoid selecting 'Next treatment' as this immediately puts the Accura into self-test mode. 2) After 15 minute, turn the power on. 3) After successful completion of self-test, resume therapy per the Accura Operator's Manual. 4) If the instrument still does not complete a successful self-test, additional time in 'stand-by' mode may be required. Repeat steps 1-3. 5) If repeated attempts do not yield a successful calibration test, then remove the instrument from service and contact Baxter Global Technical Services. Any questions were referred to the Center for One Baxter at 1-800-422-9837.
Initiated
2005-07-13
Root cause
Device Design
Terminated
2008-10-29
Distribution
Nationwide and internationally to China and Belgium
Recall number
Z-1380-05
Recall status
Terminated
CDRH record ID
40464
FDA classified
2005-08-18
Recall event number
32631

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗