Open original FDA record ↗Recall record
- FDA identifier
- Z-1380-05
- Company in FDA record
- Baxter Healthcare Renal Div
- FDA classification date
- Aug 18, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
In the event therapy is terminated, as part of routine instrument programming or therapy maintenance, the Accura instrument may experience difficulty in performing a successful self-test (failure) upon an immediate attempt to restart therapy
Source facts
- Action
- Baxter sent Important Product Information letters dated 7/13/05 to all Accura device customers, to the attention of the Director of Nursing, via first class mail. The letter informed the accounts of the on-going investigation into the potential self-test failure associated with the Accura instrument, and provided the customers with steps to follow if they encounter failed instrument self-tests. These guidance steps include: 1) Place the instrument into 'stand-by' mode. Generally, 15 minutes in 'stand-by' mode is sufficient to allow successful instrument self-test procedures. Enter 'stand-by' mode by turning off the power by using the upper power switch, or by selecting 'Accura off' from the user interface. Avoid selecting 'Next treatment' as this immediately puts the Accura into self-test mode. 2) After 15 minute, turn the power on. 3) After successful completion of self-test, resume therapy per the Accura Operator's Manual. 4) If the instrument still does not complete a successful self-test, additional time in 'stand-by' mode may be required. Repeat steps 1-3. 5) If repeated attempts do not yield a successful calibration test, then remove the instrument from service and contact Baxter Global Technical Services. Any questions were referred to the Center for One Baxter at 1-800-422-9837.
- Initiated
- 2005-07-13
- Root cause
- Device Design
- Terminated
- 2008-10-29
- Distribution
- Nationwide and internationally to China and Belgium
- Recall number
- Z-1380-05
- Recall status
- Terminated
- CDRH record ID
- 40464
- FDA classified
- 2005-08-18
- Recall event number
- 32631