Skip to content

Public FDA record

Z-1381-2023: NanoKnife 5-Probe Procedure Pack 15 cm - Activation. Indicated for the surgical ablation of soft tissue.

Recall record Z-1381-2023 for Angiodynamics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1381-2023
Company in FDA record
Angiodynamics, Inc.
FDA classification date
Apr 12, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Not programmed in accordance with specification. The programming affects the RFID function and does not allow the NanoKnife probes to be recognized by the NanoKnife generator resulting in a delay in procedure

Source facts

Action
AngioDynamics issued Urgent Medical Device Recall Letter on March 2, 2023 via Federal Express. Letter states reason for recall health risk and action to take: IMMEDIATELY Stop using the product subject to recall. Remove any affected (recalled) product from your inventory (whether in labs, Central Supply, Shipping and Receiving or ANY other location. Segregate this product in a secure location for return to AngioDynamics, Inc. Forward a copy of this recall notification to all sites to which you have distributed affected product. 2. ComplFx Reply Verification Tracking Form: Attn: NK Probe 5-Pack Recall Coordinator Fax number 1-855-273-0519 3. Package and Return the Recalled Product. Package any product that is being returned in an appropriate shipping box. Write the RMA number on the RMA/Address label (provided on the Recall Verification Tracking Form) and affix the label to the outside of the shipping box. Seal the box and return to: AngioDynamics, Inc. 24 Native Drive Queensbury, NY 12804 Attn: NK Probe 5-Pack Recall Coordinator
Initiated
2023-03-02
Root cause
Under Investigation by firm
Distribution
US Nationwide Distribution: CA,CO, FL, IA, NY, OH
Recall number
Z-1381-2023
Recall status
Open, Classified
CDRH record ID
199181
FDA classified
2023-04-12
Recall event number
91909

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗