Skip to content

Public FDA record

Z-1381-2024: Medivance Neonatal ArcticGel Pads, REF: 31802, 3180202, 31802C

Recall record Z-1381-2024 for Medivance Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1381-2024
Company in FDA record
Medivance Inc.
FDA classification date
Mar 26, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Neonatal pads are experiencing reduced water flow, which may reduce or prevent water heating or cooling performance of the pads that are part of a temperature management system.

Source facts

Action
On 2/16/24, BD mailed or emailed recall notices to customers who were asked to do the following: 1) Immediately Discontinue Use 2) Immediately quarantine affected pads, record your inventory totals on the Customer Response Form, discard all devices within your facility's control per your facility's procedures. 3) Share this notice with any users of the product within your facilities or with any facility affected product was distributed to. 4) Complete and return the attached Customer Response Form to the BD contact noted on the form. If you require further assistance contact BD Customer Support at 1-844-823-5433. Say "Recall" when prompted M-F 8am -5pm CT.
Initiated
2024-02-16
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution including in the states of IL, AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Belgium, Australia, Singapore, South Korea, Taiwan, Japan, China, Thailand.
Recall number
Z-1381-2024
Recall status
Open, Classified
CDRH record ID
206166
FDA classified
2024-03-26
Recall event number
94030

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗