Z-1382-2022: Vivo 45LS Ventilator- intended to provide continuous or intermittent ventilatory support for the care of

Recall record Z-1382-2022 for Breas Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1382-2022
Company in FDA record
Breas Medical, Inc.
FDA record creation date
Jul 13, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A forced shutdown of one of the processors did not generate a watch dog alarm, would develop into a failure AND the instructions to monitor a ventilator dependent patient are NOT followed, the health consequences could potentially be Permanent impairment or life threatening if medical intervention is not obtained

Source facts

Action
Breas Medical contacted consignees initially by phone beginning 6/2/22 followed by email Urgent Medical Device Recall Firmware Upgrade letter Letter states reason for recall, health risk and action to take: Customers are required to identify all affected Vivo 45 LS ventilators and apply the firmware upgrade to firmware version 5.0.7 (or later) within 12 months. Breas has implemented a firmware correction to resolve the problem and can confirm that Vivo 45LS with firmware versions 5.0.7 or higher, distributed after March 29th, 2022 are not affected by this issue. Confirmation of receipt We kindly ask you to acknowledge receipt of this Recall Notice together with your choice to remedy the issue using attached reply form Vivo 45 LS Firmware Upgrade Customer Reply Form. Helpline +1 (855) 436-8724
Initiated
2022-06-02
Root cause
Device Design
Distribution
Nationwide
Recall number
Z-1382-2022
Recall status
Open, Classified
CDRH record ID
194408
FDA record created
2022-07-13
Recall event number
90433

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗