Open original FDA record ↗Recall record
- FDA identifier
- Z-1383-2012
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Apr 9, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
During regular product monitoring, Siemens became aware of a potential issue with the AXIOM Luminos dRF. Firm became aware of a potential issue with torque wrenches - during a regular scheduled calibration of the wrench, it was discovered that the wrench was out of the specification. The wrench was over-torquing approximately 21% for low-end readings (50 ftlbs) and 9% for high-end readings (250
Source facts
- Action
- Siemens initiated a recall/correction on February 16, 2012 by issuing Update Instruction XP056/11/S to the affected customers. All screws on the unit support on AXIOM Luminos dRF units will be replaced by a Siemens service engineer. For information regarding this recall call 610-219-4834.
- Initiated
- 2012-02-16
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-11-06
- Distribution
- Nationwide Distribution including California
- Recall number
- Z-1383-2012
- Recall status
- Terminated
- CDRH record ID
- 108199
- FDA classified
- 2012-04-09
- Recall event number
- 61483