Z-1383-2026: MONARCH Platform, Bronchoscopy, Catalog Numbers: MON-000005-01 , MON-000005-01R , MON-000006 , MON-000006

Recall record Z-1383-2026 for Auris Health, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1383-2026
Company in FDA record
Auris Health, Inc
FDA classification date
Feb 17, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Device for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left position; and if right position was previously selected, this may lead to unanticipated robotic arm positioning toward patient-left, may result in contact with patient/endotracheal tube.

Source facts

Action
On 1/21/2026, correction notices were emailed to customers informing them of the following: Should a MONARCH application restart occur during procedure setup, please clear the fault message and then perform a full system restart. A full system restart will prevent this issue from occurring as it will require the user to perform patient-side selection and alignment again. Once the software update to address the issue is available, the firm will contact customers to schedule installation. Complete and return the acknowledgement form via email to MonarchSupport@its.jnj.com This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potential affected devices have been transferred. Please transfer this notice to other organizations on which this action has an impact. Maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective actions. If you have questions contact the firm at +1 (800) 434-0032 (toll-free within the United States), +1 (650) 264-6000 (Worldwide) or via email at MonarchSupport@its.jnj.com For Medical Information requests, visit firm's website at https://www.jnjmedtech.com/mir.
Initiated
2026-01-21
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution in the states of MN, FL, NJ, NY, VA, WI, AZ, AL, MS, LA, MI, MA, MT, PA, WV, GA, DE, TX, CA, OH, IN, SC, CT, ND, MO, DC, IL, NE, ID, WA, NC, TN, OR, CO, SD, NM, VT, MD and the countries of Hong Kong, China, Canada.
Recall number
Z-1383-2026
Recall status
Open, Classified
CDRH record ID
218085
FDA classified
2026-02-17
Recall event number
98270

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗