Open original FDA record ↗Recall record
- FDA identifier
- Z-1384-2009
- Company in FDA record
- Stryker Howmedica Osteonics Corp.
- FDA classification date
- Jul 22, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Product deviation : Specific lots of sprayed HA hip stems did not meet Stryker's Internal Material Specification for tensile bond strength and crystallinity.
Source facts
- Action
- An "Urgent Product Recall" letter was sent via Federal Express on January 6, 2009 with return receipt to Stryker Branches/Agencies, Hospital Risk Management, Hospital Chief of Orthopaedics and Surgeons. The letter described the affected product, issue, potential hazards and risks. Stryker Branches and Hospital Risk Management were asked to examine their inventory for affected product. Stryker Representatives were instructed to contact customers to reconcile and return recalled product to Stryker Orthopaedics. All letters include a response form to be completed and returned indicating that they have been informed of the recall and the amount of product on hand, if any. Direct questions about the recall to Stryker Orthopaedics at 1-201-831-5718.
- Initiated
- 2009-01-06
- Root cause
- Process control
- Terminated
- 2012-06-25
- Distribution
- Nationwide Distribution.
- Recall number
- Z-1384-2009
- Recall status
- Terminated
- CDRH record ID
- 78803
- FDA classified
- 2009-07-22
- Recall event number
- 50901