Z-1384-2012: SenDx Medical Inc ABL80FLEX Blood Gas, pH, Electrolyte, and Metabolite System The ABL80 FLEX i

Recall record Z-1384-2012 for Sendx Medical Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1384-2012
Company in FDA record
Sendx Medical Inc
FDA classification date
Apr 9, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The product can cause the analyzer to report values outside the published performance specifications for pH, pCO2, cNa+, cK+, cCa2+ and cCl-.

Source facts

Action
SenDx Medical sent Customers notification letters dated January 30th, 2012, to all affected customers. The letter identified the problem the product and the action needed to be taken by the customer. We have recently become aware of a possible condition in a limited number of ABLBO FLEX analyzers configured with a specific Control PCe lerinted Circuit Board), which can cause the analyzer to report values outside the published performance specifications for pH, pCOz, cNa*, cK*, cCa2* and cCl-. Customers were instructed to immediately implement Alternative 1 or 2 until the signal data PCB is replaced. Alterative I: Stop the use of all affected analyzers until corrective actions can be performed on these analyzers. Alternative 2: Verify proper function of the affected analyzers immediately following every patient sample measurement by measuring a manual QC sample. Customers were asked to complete the fax form acknowledging receipt of the notice and provided a contact for Technical Support at 1-800-736-0600 opt.4.
Initiated
2012-01-30
Root cause
Nonconforming Material/Component
Terminated
2012-07-13
Distribution
One US consignee in state of OH. Other consignees are international such as Australia, France, Netherlands, Denmark, China, United Kingdom, Turkey, Switzerland, Japan, India, Poland, Spain, and Germany. Distribution list submitted with supporting doc.s
Recall number
Z-1384-2012
Recall status
Terminated
CDRH record ID
107472
FDA classified
2012-04-09
Recall event number
61178

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗