Open original FDA record ↗Recall record
- FDA identifier
- Z-1384-2022
- Company in FDA record
- Agilent Technologies, Inc.
- FDA record creation date
- Jul 14, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
Label provided for substrate vail contained incorrect expiration date.
Source facts
- Action
- On 05/20/2022 and 05/27/2022, Agilent Technologies, Inc. had conference calls with the two Medical Laboratory Facilities that were distributed the affected product to inform them of the situation of the incorrect expiration date on the substrate vials (incorrect expiration dated of 6-30-22. The correct expiration date should be 4-30-22 Customer are instructed/informed: - To immediately cease using Anti-FITC-AP CISH Accessory Kit Lot Numbers 06586309 and 06646626, and to return any remaining labels to Agilent, and return any unused kits to Agilent. - That Agilent is completing production of a new lot of the Kit, and will be shipping to customers on Monday, 05/23/2022, to replace those expired kits. Kits will be provided Free of Charge. - To review test data/results from 05/01/2022 onwards as appropriate in accordance with their applicable procedures. On 06/10/2022, an "Urgent Product Removal" letter was emailed to customer informing them of the removal action. For any questions or further assistance, contact Agilent's Field Action Team at fieldactions@agilent.com
- Initiated
- 2022-05-20
- Root cause
- Labeling Change Control
- Distribution
- U.S. Nationwide distribution in the states of NY and VA.
- Recall number
- Z-1384-2022
- Recall status
- Open, Classified
- CDRH record ID
- 194410
- FDA record created
- 2022-07-14
- Recall event number
- 90435