Z-1384-2022: Kit Label: Anti-FITC-AP CISH Accessory Kit (Dako Omnis), Box A REF K589911-2 Anti-FITC-AP CISH Accessory

Recall record Z-1384-2022 for Agilent Technologies, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1384-2022
Company in FDA record
Agilent Technologies, Inc.
FDA record creation date
Jul 14, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Label provided for substrate vail contained incorrect expiration date.

Source facts

Action
On 05/20/2022 and 05/27/2022, Agilent Technologies, Inc. had conference calls with the two Medical Laboratory Facilities that were distributed the affected product to inform them of the situation of the incorrect expiration date on the substrate vials (incorrect expiration dated of 6-30-22. The correct expiration date should be 4-30-22 Customer are instructed/informed: - To immediately cease using Anti-FITC-AP CISH Accessory Kit Lot Numbers 06586309 and 06646626, and to return any remaining labels to Agilent, and return any unused kits to Agilent. - That Agilent is completing production of a new lot of the Kit, and will be shipping to customers on Monday, 05/23/2022, to replace those expired kits. Kits will be provided Free of Charge. - To review test data/results from 05/01/2022 onwards as appropriate in accordance with their applicable procedures. On 06/10/2022, an "Urgent Product Removal" letter was emailed to customer informing them of the removal action. For any questions or further assistance, contact Agilent's Field Action Team at fieldactions@agilent.com
Initiated
2022-05-20
Root cause
Labeling Change Control
Distribution
U.S. Nationwide distribution in the states of NY and VA.
Recall number
Z-1384-2022
Recall status
Open, Classified
CDRH record ID
194410
FDA record created
2022-07-14
Recall event number
90435

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗