Patella drills are marked with incorrect diameter size of 5.6mm, when they should be 5.4mm. Porous patella overdrills are marked with incorrect diameter size of 5.4 mm, when they should be 5.6 mm. If the incorrect drill is selected during surgery, there is risk of the implant loosening, implant not properly fitting, and/or fractured bone.
Source facts
Action
On 2/28/23, field safety notices were emailed to surgical sales agents who were asked to document the quantity of affected devices complete and return the acknowledgement form. Customer service contact, customerservice@enovis.com, was provided to report failures or to request replacements. Customers with questions can phone (914) 450-3931, or email livia.kuruvila@enovis.com.
Initiated
2023-02-28
Root cause
Process change control
Distribution
US: TX, IN, MD, CA, KS, AL, HI, WA
Recall number
Z-1384-2023
Recall status
Open, Classified
CDRH record ID
199178
FDA classified
2023-04-12
Recall event number
91906