Skip to content

Public FDA record

Z-1385-2019: enFlow Disposable Cartridge, used for intravenous warming therapy with fluid and blood solutions.

Recall record Z-1385-2019 for Vyaire Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1385-2019
Company in FDA record
Vyaire Medical
FDA classification date
Jun 26, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Testing has demonstrated aluminum elution from the enFlow Disposable Cartridge during fluid warming.

Source facts

Action
Vyaire Medical notified consignees on about 03/13/2019 via URGENT NOTIFICATION OF RECALL letter. Consignees were instructed to conduct a 100% physical inventory to identify any affected product on hand, destroy all affected product in-stock in accordance with each facility's destruction protocol and if a facility is unable to destroy the product on site, or requires assistance, they were instructed to contact VyaireSupport@stericycle.com or call for assistance. Customers were also instructed to complete and return the Customer Response Form to VyaireSupport@stericycle.com, report any adverse reactions or quality problems to Vyaire's Technical Support Department by e-mail GMB-RS-DISP-Complaint-Intake@carefusion.com or telephone at 1-800-231-2466 prompt #1. If customers further distributed the product, they were instructed to forward a copy of the URGENT NOTIFICATION OF RECALL letter and Response Form to all customers and include the contact information to Vyaire for tracking purposes.
Initiated
2019-03-13
Root cause
Device Design
Terminated
2020-09-22
Distribution
Worldwide Distribution. US nationwide, Brazil, Canada, Chile, Colombia, Ecuador, Hong Kong, Hungary, Israel, India, Jordan, Korea, Kuwait, Kazakhstan, Lebanon, Macedonia, Mexico, Malaysia , Myanmar, Netherlands, New Zealand, Oman, Peru, Qatar, Russian Federation, Saudi Arabia, Thailand, Turkey, United Republic of Tanzania, South Africa, Panama, and Viet Nam.
Recall number
Z-1385-2019
Recall status
Terminated
CDRH record ID
171533
FDA classified
2019-06-26
Recall event number
82393

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗