Open original FDA record ↗Recall record
- FDA identifier
- Z-1385-2022
- Company in FDA record
- Hamilton Medical AG
- FDA record creation date
- Jul 14, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
When starting a humidifier, intended for respiratory gas conditioning during invasive and noninvasive mechanical ventilation, with an empty chamber, and water is not filled into the chamber, the low water level alarm will be delayed and medical staff are unaware that there is no water in the chamber.
Source facts
- Action
- On 01/05/21, Field Safety Corrective Action notices were sent to customers who were asked to do the following: Follow the instructions in the HAMILTON-H900 Operator s Manual 624431 section 3.4. Before connecting a new patient to the humidifier; " Verify that the alarms operate correctly. " Ensure that the water supply to the humidifier chamber is functioning properly. Distribute this notification to all end users within your organization. A Medical Representative will be in contact to schedule service with a software update. Customers with additional questions or concerns were encouraged to call Technical Support at 1-800-426-6331, option #2.
- Initiated
- 2021-01-05
- Root cause
- Employee error
- Distribution
- US: WI, FL, CA, NV, PA, TX, PA
- Recall number
- Z-1385-2022
- Recall status
- Completed
- CDRH record ID
- 194457
- FDA record created
- 2022-07-14
- Recall event number
- 90460