Open original FDA record ↗Recall record
- FDA identifier
- Z-1386-2012
- Company in FDA record
- Roche Diagnostics Operations, Inc.
- FDA classification date
- Apr 9, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Accu-Chek Aviva Nano Blood Glucose meters that were manufactured for the purpose of validating a manufacturing process for product to be distributed outside the US were distributed in the US through eBay by an unauthorized third party. These meters were not intended for distribution, were not verified to meet final product specifications and were not distributed to customers with the necessary pr
Source facts
- Action
- Roche sent an "URGENT MEDICAL DEVICE REMOVAL" letter dated March 14, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers were instructed to immediately return their ACCU-CHEK Aviva Nano Meter to the firm in an enclosed pre-paid envelope via USPS. Contact Roche Product Removal Services at 1-888-345-5352 between 8:00 am and 5:00 pm for questions regarding this notice.
- Initiated
- 2012-03-12
- Root cause
- Employee error
- Terminated
- 2012-11-06
- Distribution
- Nationwide Distribution-USA (nationwide) including the states of AZ, NY, RI, TN, NJ, IN, TX, AL, NV, CA, OR, KY, IL, AR, NM, NE, MD, MO, GA, and SD.
- Recall number
- Z-1386-2012
- Recall status
- Terminated
- CDRH record ID
- 108011
- FDA classified
- 2012-04-09
- Recall event number
- 61405