Maquet has received 3 confirmed complaints related to the Li-lon Battery used with the Maquet CARDIOSAVE Hybrid IABP and Maquet CARDIOSAVE Rescue IABP of the battery venting after being accidentally dropped.
Source facts
Action
Maquet contacted their customers via letter on 12/23/16 and additional customers identified were notified on 1/10/17 and issued a recall letter/return response form. The letter identified the affected product problem and actions to be taken. Customers were instructed to complete and return the response form via fax to 1-973-629-1518 or email to cardiosavebattery@getinge.com
Initiated
2016-12-23
Root cause
Device Design
Terminated
2019-03-01
Distribution
Worldwide Distribution - US Nationwide
Recall number
Z-1386-2017
Recall status
Terminated
CDRH record ID
152101
FDA record created
2017-03-02
Recall event number
76052