Z-1386-2025: Access 2 Immunoassay Analyzer, Catalog Number 81600N

Recall record Z-1386-2025 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1386-2025
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Mar 17, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Beckman Coulter identified that an issue for Access 2 Sample Motors (B48016) due to improper alignment, causing the motors to rub against the sample wheelbase plate and index sensor. The issue was traced to a worn tool used in production (at the supplier), affecting Access 2 sample motors. This could lead to a potential delay in reporting patient test results.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notification letter dated 1/31/25 was sent to customers. If the error occurs: 1. Confirm the system is in the Ready or Not Ready mode. 2. Go to the Initialize System window. To get to this window from the Main Menu, select Diagnostics F7 to display the Diagnostics menu, then select Initialize System F1. The Initialize System window is displayed with a confirmation message. 3. Select Yes F1 to initialize the instrument. Contact your Beckman Coulter representative if your laboratory has experienced sample carousel motion errors that are not resolved by initializing the analyzer. Your Beckman Coulter service representative will replace the sample motor during your next semi-annual preventative maintenance or repair visit. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: Electronically, if you received this communication via email. Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact our Customer Support: From our website: http://www.beckmancoulter.com
Initiated
2025-02-05
Root cause
Nonconforming Material/Component
Distribution
Worldwide - US Nationwide distribution.
Recall number
Z-1386-2025
Recall status
Open, Classified
CDRH record ID
212416
FDA classified
2025-03-17
Recall event number
96248

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗