Open original FDA record ↗Recall record
- FDA identifier
- Z-1387-06
- Company in FDA record
- Radiometer America Inc
- FDA classification date
- Aug 10, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Under certain circumstances, the ABL 700 &800 Series Blood Gas Analyzers may experience leakage current into the measuring system. Consequently, the analyzers intermittently provide incorrect (too low) result values for Calcium (Ca) and Sodium (Na).
Source facts
- Action
- The consignees were notified of this recall by a letter, dated 9/29/2005. The letter informs the customers of this potential problem and the availability of new updated software which will correct the problem. The letter goes on to inform the customer that a Radiometer Service Engineer wil be contacting them shortly in order to arrange for a convenient time to install the software update. Along with the letter, the customer is provided a copy of an additional page and a CD concerning the function and operation of the new software, which is to be added to the customer''s Operator''s Manual.
- Initiated
- 2005-09-29
- Root cause
- Other
- Terminated
- 2012-04-17
- Distribution
- Nationwide.
- Recall number
- Z-1387-06
- Recall status
- Terminated
- CDRH record ID
- 45026
- FDA classified
- 2006-08-10
- Recall event number
- 34963