Skip to content

Public FDA record

Z-1387-2016: SYNTHECEL Dura Repair, 510K # K131792, K113071, Class II, Classification name: Dura Substitute Device lis

Recall record Z-1387-2016 for Synthes (USA) Products LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1387-2016
Company in FDA record
Synthes (USA) Products LLC
FDA record creation date
Oct 11, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was reported that the SYNTHECEL Dura Repair could adhere to various anatomical structures post implantation, which may expose patients to potential risks in the event a surgeon intends to temporarily apply and subsequently remove a SYNTHECEL implant. SYNTHECEL Dura Repair labeling and promotional material are being updated to clarify that SYNTHECEL is intended to remain in place and is not intended to be removed.

Source facts

Action
Siemens mailed out an Urgent: Field Safety Notification letter informing customers to review the replacement package insert (GP2715-C) and discard the outdated package insert (GP2715-B) and outdated promotional materials. Customers were asked to complete the verification section (page 3 of the recall notification) and send a copy of the completed Verification Section by Fax (888)912-7351 or Scan/Email to Synthes5875@stericycle.com
Initiated
2015-12-08
Root cause
Labeling False and Misleading
Terminated
2016-08-26
Distribution
Nationwide: NY, PA, AL, WI, FL, MO, TN, OH, MI, MN, CO, CA, NV, AZ, LA, OR, TX, IA, IN, MS, IL, AR, WV, HI, VA, MD, WA, OK, MA, KY, NJ, CT, No foreign accounts.
Recall number
Z-1387-2016
Recall status
Terminated
CDRH record ID
144463
FDA record created
2018-10-11
Recall event number
73559

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗