Skip to content

Public FDA record

Z-1387-2023: Step Bladeless Trocars, 10 mm cannula and dilator with radially expandable sleeve and 5 - 10 mm reducer,

Recall record Z-1387-2023 for Covidien, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1387-2023
Company in FDA record
Covidien, LP
FDA classification date
Apr 12, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Medtronic was made aware of a supplier calibration issue during the contract sterilization process that may lead to unsterile product. Use of affected product may lead to an unspecified infection or, in the case of sutures used in the eye, pose a loss of vision risk.

Source facts

Action
An "URGENT PRODUCT RECALL" letter dated 3/1/23 was sent to customers. Actions to be Taken: " Immediately identify and quarantine all unused affected product (Refer to Attachment A for identifying affected product). " Please complete and return the enclosed Customer Confirmation Form, even if you have no product to return and email to rs.gmbfcamitg@medtronic.com. " Product purchased directly from Medtronic please contact rs.covidienfeedbackcustomerservice@medtronic.com for Return Goods Authorization (RGA). Your local Medtronic Representative can assist you as necessary in initiating the return of this product. " Credit for the returned affected product will be issued based on the RGA number. " Return all unused affected product(s) in your inventory to: Medtronic, Attn: Field Returns Dept. 195 McDermott Road, North Haven, CT 06473 USA " If purchased from a distributor, contact your distributor directly to arrange for the return of the product back to your distributor. " Forward this notification to any entities within your organization or to any other organization where the potentially affected product has been transferred or further distributed. Local contact details: Adverse reactions or quality problems experienced with this product should be reported to FDA and Medtronic: " Online at http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail), or Call FDA (800) FDA-1088 " Contact Medtronic Customer Service Support at rs.covidienfeedbackcustomerservice@medtronic.com or call 800-962-9888, option 2. We regret any inconvenience this may cause. We are committed to patient safety and appreciate your prompt attention to this matter. If you have any questions regarding this communication, please contact your Medtronic Representative Customer Service at 800-962-9888, option 2.
Initiated
2023-03-01
Root cause
Process control
Distribution
Worldwide Distribution: US (nationwide) to States of: AR, CA, FL, GA, LA, MA, NC, NY, OH, WA; and countries of: Canada, Colombia, El Salvador, Indonesia, Nepal, Puerto Rico, Taiwan, and Thailand.
Recall number
Z-1387-2023
Recall status
Open, Classified
CDRH record ID
199083
FDA classified
2023-04-12
Recall event number
91873

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.