Z-1387-2024: CareEvent Event Management System, Model Numbers 866435 and 866436 (CareEvent Upgrade), Software Version

Recall record Z-1387-2024 for Philips North America. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1387-2024
Company in FDA record
Philips North America
FDA classification date
Mar 27, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was found that the Care Assist mobile application (version 4.1.1 and earlier) would not play custom ringtones when iOS device screens were locked with a passcode.

Source facts

Action
An URGENT Medical Device Correction dated 2/27/24 was sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients or users: 1. The Care Assist mobile application should be used in conjunction with other monitoring devices such as a bedside monitor, telemetry monitor and central monitor which continue to provide visual and audible alarm information to inform clinicians of changes in patient status. 2. Customers should update their iOS devices to version 4.1.2 of Care Assist mobile application or greater to eliminate the software defect affecting the annunciation of custom ringtones. 3. Customers should complete the Urgent Medical Device Correction Response Form at the end of the notification to submit both their acknowledgement of this recall and confirm understanding of actions to be taken. 4. This communication should be shared with all clinical staff to review and understand. 5. Place this Urgent Medical Device Correction notification with the documentation of the Care Event and Patient Information Center iX (PlC ix) Revision 4. For Distributors: Philips is requesting customers to return a Response Form to acknowledge receipt and understanding of the Medical Device Correction and confirm that the information from this Letter has been properly distributed to all users that handle the affected product. Together with this letter we are providing a list of affected products that Philips has sold to your organization. As distributor of the affected products, we kindly request that you: Send the attached Medical Device Correction to each customer to whom you have distributed any affected product as soon as possible and no later than? days, together with the Reply Card. Perform a good faith effort to ensure Notification and Reply Form arc deliver to the customer with a minimum of three attempts, and if possible, using multiple contact methods. Actions planned by Philips to correct the problem 1, The action by Philips
Initiated
2024-02-28
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution.
Recall number
Z-1387-2024
Recall status
Open, Classified
CDRH record ID
206537
FDA classified
2024-03-27
Recall event number
94185

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗