Open original FDA record ↗Recall record
- FDA identifier
- Z-1388-2025
- Company in FDA record
- Beckman Coulter, Inc.
- FDA classification date
- Mar 17, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
the DxI 9000 encoder wheel was misaligned on the motor shaft, leading to increased slippage errors and causing the instrument to stop, posing a high safety risk due to potential delays in patient results.
Source facts
- Action
- An URGENT MEDICAL DEVICE RECALL notification letter dated 1/30/25 was sent to customers. Affected DxI 9000 Access Immunoassay Analyzers could experience an increase in the following errors: IncubateWheel.Axis MachineControlStatus(Encoder reports that axis position is out of tolerance) WashWheel.Axis MachineControlStatus(Encoder reports that axis position is out of tolerance) This can cause the analyzer to enter the red, or Stopped, system status and any in-process tests are cancelled. There could be a delay in reporting patient test results to the physician. If the error occurs: 1. Confirm that the DxI 9000 Access Immunoassay Analyzer is in either the Paused or Stopped system state. 2. Select the System state button in the lower left corner of the user interface. 3. Select Initialize. When initialization is complete, the analyzer returns to the green, or Running, state. If your laboratory has experienced the issue and it is not resolved by initializing the analyzer, contact your Beckman Coulter representative and schedule a service visit. Beckman Coulter has identified the root cause of this issue and is no longer distributing affected motors for the Incubate and Wash Wheel assemblies. A Beckman Coulter service representative will schedule a visit to replace any affected motors on your DxI 9000 Access Immunoassay Analyzer. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact our Customer Support Center; From our website: http://www.beckmancoulter.com
- Initiated
- 2025-02-05
- Root cause
- Process design
- Distribution
- US Nationwide. Global Distribution.
- Recall number
- Z-1388-2025
- Recall status
- Open, Classified
- CDRH record ID
- 212435
- FDA classified
- 2025-03-17
- Recall event number
- 96256