Z-1388-2025: DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137.

Recall record Z-1388-2025 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1388-2025
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Mar 17, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

the DxI 9000 encoder wheel was misaligned on the motor shaft, leading to increased slippage errors and causing the instrument to stop, posing a high safety risk due to potential delays in patient results.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notification letter dated 1/30/25 was sent to customers. Affected DxI 9000 Access Immunoassay Analyzers could experience an increase in the following errors: IncubateWheel.Axis MachineControlStatus(Encoder reports that axis position is out of tolerance) WashWheel.Axis MachineControlStatus(Encoder reports that axis position is out of tolerance) This can cause the analyzer to enter the red, or Stopped, system status and any in-process tests are cancelled. There could be a delay in reporting patient test results to the physician. If the error occurs: 1. Confirm that the DxI 9000 Access Immunoassay Analyzer is in either the Paused or Stopped system state. 2. Select the System state button in the lower left corner of the user interface. 3. Select Initialize. When initialization is complete, the analyzer returns to the green, or Running, state. If your laboratory has experienced the issue and it is not resolved by initializing the analyzer, contact your Beckman Coulter representative and schedule a service visit. Beckman Coulter has identified the root cause of this issue and is no longer distributing affected motors for the Incubate and Wash Wheel assemblies. A Beckman Coulter service representative will schedule a visit to replace any affected motors on your DxI 9000 Access Immunoassay Analyzer. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact our Customer Support Center; From our website: http://www.beckmancoulter.com
Initiated
2025-02-05
Root cause
Process design
Distribution
US Nationwide. Global Distribution.
Recall number
Z-1388-2025
Recall status
Open, Classified
CDRH record ID
212435
FDA classified
2025-03-17
Recall event number
96256

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗