Skip to content

Public FDA record

Z-1389-2009: Welch Allyn AED 10 automated external defibrillator and MRL Jumpstart (collectively "AED 10")

Recall record Z-1389-2009 for Welch Allyn Protocol, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1389-2009
Company in FDA record
Welch Allyn Protocol, Inc
FDA classification date
Jun 23, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Issue #1: AED 10 displayed a "Do Not Use" symbol as a result of a blown F1 fuse. Issue #2: AED 10 models that incorporate a spring connector with the speaker experienced intermittent electrical connection to the circuit board causing a loss of voice prompts.

Source facts

Action
Welch Allyn sent an Urgent Medical Device Recall letter dated February 26, 2009 to all distributors and customers, describing 5 failure modes. Customers were instructed to respond to the letter by calling 888-345-5356, and decide on replacement or discounted new purchase.
Initiated
2009-02-26
Root cause
Component design/selection
Terminated
2012-03-07
Distribution
Worldwide distribution.
Recall number
Z-1389-2009
Recall status
Terminated
CDRH record ID
80550
FDA classified
2009-06-23
Recall event number
51496

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗