Software issue with compounding pump resulting in over delivery of component ingredients during routine operations.
Source facts
Action
Baxa sent a Safety Alert letter dated March 23, 2009 to all affected consignees. The letter identified the affected product, problem and recommended actions with specific instructions related to the issue. Additionally, the firm will be providing a warning label for the affected product as soon as possible and is currently working on a product upgrade to correct the issue. For assistance contact Baxa Technical Support at 800-678-2292. The firm is also providing a 24 hour a day, 7 days a week Technical Support for the affected product.
Initiated
2009-03-23
Root cause
Software design
Terminated
2012-04-10
Distribution
Worldwide Distribution - USA Nationwide including Canada.
Recall number
Z-1389-2012
Recall status
Terminated
CDRH record ID
81930
FDA classified
2012-04-09
Recall event number
52024