Z-1389-2022: GE Healthcare Aisys CS2 Anesthesia System, Part number 1011-9050-000; and Avance CS2 and Avance CS2 Pro A

Recall record Z-1389-2022 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1389-2022
Company in FDA record
GE Healthcare, LLC
FDA classification date
Jul 19, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

There is a potential reversal of the O2 and air cylinder pressure transducer connections.

Source facts

Action
The recalling firm issued letters beginning 6/1/2022 via traceable means to the Chief of Anesthesia, Director of Biomedical/Clinical Engineering, and Health Care Administrator/Risk Manager at their consignee locations. The letter was flagged that consignees should ensure all potential users in their facility are made aware of this safety notification and the recommended actions. The letter provided the affected products, safety issue, and actions to be taken. The customer was informed they could continue to use the anesthesia system by following the instructions provided in the letter prior to use. A Medical Device Notification Acknowledgement Response form was enclosed to acknowledge receipt and understanding of the notice. The consignee was to record their anesthesia system serial number, whether it passed or failed the test, and the date of the test. The response form was supposed to be completed and returned no later than 30 days from receipt.
Initiated
2022-06-01
Root cause
Process control
Terminated
2024-10-21
Distribution
Worldwide Distribution. US nationwide, Australia, Austria, Belgium, Bolivia, Brazil, Canada, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Italy, Japan, Kenya, Korea, Kuwait, Malaysia, Mexico, Montenegro, Nepal, Netherlands, New Zealand, Norway, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and Uruguay.
Recall number
Z-1389-2022
Recall status
Terminated
CDRH record ID
194361
FDA classified
2022-07-19
Recall event number
90424

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗