Open original FDA record ↗Recall record
- FDA identifier
- Z-1391-2009
- Company in FDA record
- Biosite Inc
- FDA classification date
- May 8, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The recall was initiated because the use of the affected lot may lead to reporting falsely low values for troponin I and myoglobin, possibly resulting in missed or incorrect diagnosis.
Source facts
- Action
- The recall was initiated with "Urgent Medical Device Recall" notices, dated 4/28/2009, with attached "Customer/Distributor Verification Form" being sent to the affected customers via fax, e-mail or direct mail on 04/28/2009. Customers were informed of the reason for the recall and instructed to immediately discontinue all use of the listed products and discard all affected product in accordance with your their regulations. The customers were give two options; "Option 1: Complete and FAX the enclosed Verification Form within 10 days to confirm your receipt of this notice, to indicate the number of kits discarded from your inventory and product replacement will be sent to you. Option 2: Contact inverness Medical Technical Service at 1-877-441-7440 and select option 2 from the phone prompt. A representative will confirm the number of kits discarded from your inventory and arrange for replacement product to be sent to you." Customers were also provided telephone, fax and e-mail contact information for Biosite should they have any questions about the information contained in the recall notification.
- Initiated
- 2009-04-27
- Root cause
- Process control
- Terminated
- 2010-07-06
- Distribution
- Nationwide Distribution
- Recall number
- Z-1391-2009
- Recall status
- Terminated
- CDRH record ID
- 81652
- FDA classified
- 2009-05-08
- Recall event number
- 51925