Open original FDA record ↗Recall record
- FDA identifier
- Z-1391-2022
- Company in FDA record
- Aesculap Implant Systems LLC
- FDA record creation date
- Jul 15, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Nonfunctional key due to the potential for damage to the key pins cause by force of use during assembly of the tibia locking ring.
Source facts
- Action
- A notification letter was issued November 12, 2019 mailed USPS certified with registered return receipt. The letter identified item number NP462R EnduRo Special Key for Tibia Locking Ring F1 10/12mm as being removed from the market for a new version NP462RM after overall system modification. Users are instructed to 1) Review inventory and check that only the latest version (NP462RM and NP454R) are being used. 2) Contact customer service to return any affected inventory. 3) Complete and return the acknowledgement form. Questions are to be directed to Josie McBride at (610) 984-9290.
- Initiated
- 2019-11-12
- Root cause
- Device Design
- Distribution
- US Nationwide distribution in the states of AL, AZ, CA, CO, ID, IL, LA, MA, MO, NC, OK, TX, WI.
- Recall number
- Z-1391-2022
- Recall status
- Completed
- CDRH record ID
- 194536
- FDA record created
- 2022-07-15
- Recall event number
- 90484