Skip to content

Public FDA record

Z-1391-2024: HydroFlock Sterile Flocked Collection Device, Part Number 25-3317-H.

Recall record Z-1391-2024 for Puritan Medical Products Company, Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1391-2024
Company in FDA record
Puritan Medical Products Company, Llc
FDA classification date
Mar 28, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was discovered through a customer complaint that mislabeled product was distributed. Puritan distributed product labelled as 25-3317-H lot S1014 which contained product code 25-3317-U in all packages.

Source facts

Action
URGENT: FIELD SAFETY NOTICE/MEDICAL DEVICE RECALL letters were sent to customers 21-Feb-2024. Customers were instructed to: 1. Carefully read the content of this notification. 2. Ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification. 3. Document on the Reply Form the number of the affected products currently located at your facility and confirm they have been destroyed. 4. If you have already further processed or utilized this product in a kit, assess the impact of the incorrect collection device on the final medical device / in vitro diagnostic that is distributed. 5. Notify your customers of this field action. Advise your customers to cease use or further distribution of the affected product, destroy kits containing this affected product, and re-review any results already obtained using these collection devices. Please appropriately document your notification process and let us know the end-customer feedback accordingly. 6. Attest that and that you have received and understood this notification and have notified any additional customers you may have distributed to by filling out and returning the completed enclosed Reply Form by 27-Feb-2024. Puritan Medical is coordinating with the applicable Regulatory Agencies. We fully appreciate your prompt cooperation in addressing this situation and working with us to ensure your patients are treated only with safe and effective Puritan products that you have come to rely upon. If you require additional information or on-site support, please do not hesitate to contact Puritan directly at (800) 321-2313 from Monday through Friday or by e-mail at sales@puritanmedproducts.com.
Initiated
2024-02-21
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, FL, GA, IA, ID, IL, IN, KY, MA, MI, MO, NC, NJ, NY, OH, OK, PA, TX, VA and the countries of Canada, P.R. China and Thailand.
Recall number
Z-1391-2024
Recall status
Open, Classified
CDRH record ID
206297
FDA classified
2024-03-28
Recall event number
94087

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗