Open original FDA record ↗Recall record
- FDA identifier
- Z-1392-2011
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- Feb 22, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
If the transducer is not disconnected from the system after an automatic shutdown due to thermal limits being exceeded, then the next time it is used, the thermal safety guards do not take effect. Normally, the thermal limit before automatic image freeze (and therefore power shut off) occurs at 43 degrees Centigrade.
Source facts
- Action
- Siemens sent a Customer Safety Advisory Notification letter on August 28, 2009, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed that their local Customer Service Engineer would install a software update as soon as it becomes available. Customers were instructed to pass the Customer Safety Advisory Notification letter to all who need to be aware of these issues until the corrective action is completed. For questions regarding this recall call 650-969-9112 or 800-422-8766.
- Initiated
- 2009-08-28
- Root cause
- Software design
- Terminated
- 2011-03-29
- Distribution
- Worldwide Distribution - USA (nationwide) and internationally to Canada, China, Czech Republic, Denmark, Germany, Great Britian, Hong Kong, Indonesia, Italy, Japan, Norway, Poland, Portugal, Russian Federation, Singapore, South Korea, Taiwan, and Turkmenistan.
- Recall number
- Z-1392-2011
- Recall status
- Terminated
- CDRH record ID
- 86332
- FDA classified
- 2011-02-22
- Recall event number
- 53691